Julabo LC 4 Temperature controller Bedienungsanleitung Seite 27

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 35
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 26
LC4
27
6.6. Interface parameters: r - br - P - H
RS232C
Whenever the parameter setting of the controller is not conform to
those of the connected personal computer, a modification is
necessary.
Factory setting of the RS232 interface:
Baudrate: 4800 Bauds
Parity: 2 (even)
Handshake: 1 (Hardware handshake)
Data bits: 7
Stop bits: 1
Press the cursor key until the desired parameter appears on
the MULTI-DISPLAY (LED).
Use the edit keys to set the desired parameter and
then press enter .
Adjustable interface parameters
REMOTE 0 = keypad control mode
1 = remote control mode via RS232C
BAUDRATE 12 = 1200 bauds
24 = 2400 bauds
48 = 4800 bauds
96 = 9600 bauds
PARITY 0 = no parity
1 = odd
2 = even
HANDSHAKE
0 = Protocol Xon/Xoff (software handshake)
1 = Protocol RTS/CTS (hardware handshake)
Like all parametes which can be entered through the keypad,
interface parameters are stored in memory even after the controller
is turned off.