Operating procedures
16
5.5. Applications
Directly heated bath liquid:
The bath liquid is directly heated via the heating device.
Working and safety sensors must both be immersed in the bath liquid.
Whenever countercooling is necessary (in case of temperature application near the
ambient temperature), cooling in the bath is performed through a cooling coil connected to
a solenoid valve and the MVS controller.
!
14
15
18
16
JULABO MVS
25
26
27
28
29
30
14a
15a
18a
31
16a
14 Connector: Working sensor 14a Working sensor
15 Connector: Safety sensor 15a Safety sensor
16 Connector for control cable 16a Control cable for MVS
18 Mains socket for heating device 18a Heating device / heating hood
25 MVS solenoid valve controller (order no. 9 790 000)
26 Solenoid valve 220 Volts (order no. 8 980 700)
27 Cooling coil
28 Tap water connection
29 Cooling water drain
30 Stirrer motor for bath circulation
31 Bath tank / round bottom flask with bath liquid